PORTABLE TURBIDITY METER
LaMotte
2020e Turbidimeter (Nephelometer)
The newly redesigned 2020 combines
an advanced microprocessor with a patent-pending optical chamber
resulting in higher accuracy and a wider range. The multi-detector optical configuration assures long term stability
and minimizes stray light and color interferences. All readings
are determined by the process of signal averaging over a 5 second
period. This minimizes fluctuations in readings attributed to large
particles and results in rapid, highly repeatable measurements.
Ideally suited for both low-level drinking water applications as
well as monitoring high turbidity in the field.
|
 |
 |
Patent pending optical design features focusing optics
for low level precision and accuracy. |
 |
Six user selected languages – English, Spanish,
French, Japanese, Italian, and Portuguese. |
 |
MSP430 Microcontroller used is the most advanced controller on
the market for hand held applications. |
 |
Advanced calibration algorithms. |
 |
Tube positioning ring limits vial variability. |
 |
Easy menu driven operation and large LCD display. |
 |
4000 point data log, stored results can be viewed directly on instrument
or downloaded to a computer via RS232 cable. |
| |
|
|
2020e version meets USEPA design criteria as specified
by USEPA method 180.1. |
Kits are supplied with 0, 1, and 10 NTU standard, sample bottle,
four sample tubes, and an extra battery. AC power adapter is optional.
|
Turbidity Specifications:
|
|
Unit of Measure:
|
NTU, FNU, FAU, ASBC, EBC
|
|
Range:
|
0-4000
|
|
Resolution:
|
0.01 NTU/FNU 0.00-10.99, 0.1 NTU/FNU 11.00-109.9,
1 NTU/FNU 110-4000
|
|
Accuracy:
|
±2% 0-100 NTU, ±3% above 100 NTU
|
|
Detection Limit:
|
0.05 NTU/FNU
|
|
Range Selection:
|
Automatic
|
|
Reproducibility:
|
0.02 NTU/FNU, 0.5 FAU
|
|
Light Source:
|
860nm LED (2020i ISO).